CARTO 3 Systems, Product Code M-4800-01 The intended use of the CARTO 3 System is catheter-based cardiac electrophysiological (EP) procedures.

Hazard

Some CARTO 3 Systems may get radio frequency interference through certain interconnected cable configurations resulting in jumping or disappearing catheter icon and/or error messages during ablation.

What You Should Do

Remedy

Three Urgent Customer Notification recall communications were initiated 2/9/12 with Biosense forwarding the notification to customers with affected backplane cards (electrophysiology lab managers, a notification for customers with affected back plane cards (risk managers), and a notification for customers with affected backplane cards and smarttouch module along with customer acknowledgement forms. The letters provides the customers with an explanation of the problem identified and steps to take in the event that the issue is observed. If the steps provided do not resolve the issue, customers are to contact their Biosense Webster representative. A Biosense Webster representative will contact customers to schedule an appointment for the replacement of the Backplane card. Customers were instructed to complete and return the attached Acknowledgement Form in accordance with the instructions listed on the form. If customers have questions, they can contact their BWI representative.

Products

CARTO 3 Systems, Product Code M-4800-01 The intended use of the CARTO 3 System is catheter-based cardiac electrophysiological (EP) procedures.
Brand
Biosense Webster, Inc.
UPC
Catalog # FG-5400-00  Serial numbers: 11996, 11998, 11999, 12002, 12026, 12031, 12040, 12046, 12047, 12051, 12052, 12057, 12061, 12062, 12063, 12065, 12066, 12067, 12069, 12070, 12075, 12077, 12078, 12086, 12087, 12088, 12089, 12094, 12096, 12098, 12099, and 12111.

Units Affected

134 units total (32 units in US)