DuraspanTM Dialysis Catheters, Ultra and Long Term, 15.5 Fr, Sterile EO, Manufactured for r4 Vascular, Inc . 7550 Meridan Circle North, Suite 150, Maple Grove, MN 55369. Description Model # Duraspan Kit - 24cm 62152420 Duraspan Kit - 28cm 62152820 Duraspan Kit - 32cm 62153220 Duraspan Exchange Kit - 32cm 62153228 Duraspan Kit - 36cm 62153620 Duraspan Exchange Kit - 3

Hazard

r4 Vascular, Inc. is conducting a voluntary recall of specific lots of Duraspan Dialysis Catheters. Duraspan catheters are 15.5 Fr dual lumen chronic hemodialysis catheters. Over a six month period, seven patients at five hospitals experienced separation of the luer connector from the extension tube on the Duraspan catheter. Testing during failure analysis disclosed that these lots and other lot

What You Should Do

Remedy

R4 Vascular Inc sent an Urgent Medical Device Recall letter dated May 3, 2011, to all affected customers. The letter described the problem and the product being recalled. Consignees were instructed to stop distributing the product and to quarantine the recalled lots and if they had already distributed the affected product to seek its return from their customers. A template letter to physicians and healthcare facilities was enclosed for Distributors/Retailers to send or hand deliver to their customers. For questions regarding this recall call 1-704-905-2105.

Products

DuraspanTM Dialysis Catheters, Ultra and Long Term, 15.5 Fr, Sterile EO, Manufactured for r4 Vascular, Inc . 7550 Meridan Circle North, Suite 150, Maple Grove, MN 55369. Description
Brand
R4 Vascular Inc
UPC
42 Lots Total in total: R401D035P R401D049P R401D053P R401D055P R401D056P R401D057P R401D060P R401D092P R401D097P R401D098P R401D099P R401D102P R401D123P R401D124P R401D125P R401D126P R401D169P R401D217P R401D218P R401D250P R401D251P  R402D050P R402D056P R402D057P R402D097P R402D102P R402D130P R402D138P R402D162P R402D168P R402D217P R402D250P   R403D057P R403D097P R403D123P R403D140P R403D159P R403D196P R403D217P  R404D057P R404D196P  D405D057P

Units Affected

2489 kits