Clinical Chemistry Lipase, List Numbers: 7D80-20 and 7D80-30

Hazard

Lipase results are falsely elevated when a MULTIGENT Acetaminophen test is pipetted immediately before a Lipase test from the same sample cup or tube on the ARCHITECT c8000.

What You Should Do

Remedy

The firm initiated the recall on 06/01/2007 by sending recall letters with attached Customer Reply form via FedEx to each direct account that received the recalled product. Customers were informed of the issue and given instructions for modification to the SmartWash feature in the Lipase Assay Configuration in order to prevent Acetaminophen from contaminating Lipase.

Products

Clinical Chemistry Lipase, List Numbers: 7D80-20 and 7D80-30
Brand
Abbott Laboratories Inc.
UPC
List Numbers: 7D80-20 and 7D80-30; Control Numbers: 37033HW00, 33032HW00, 30053HW00, 47048HW00, 45067HW00, 43002HW00, 42016HW00 & 42010HW00

Units Affected

7260