CyberKnife Robotic Radiosurgery System: Lung Optimized Treatment option. The CyberKnife treatment is indicated for the treatment planning and image guided stereotactic radiosurgery and precision radiotherapy of lesions, tumors and conditions anywhere in the body when radiation treatment is indicated.
Hazard
User Facility reported an anomaly where inhale and exhale CT pairs used for treatment planning did not represent the same magnitude of respiration that was displayed during treatment delivery.
What You Should Do
Accuray sent a Urgent Medical Device Correction letter dated November 7, 2011, via Federal Express to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. The information provided in this letter will be incorporated in user manuals shipped with the CyberKnife system. If you have any questions or concerns regarding this issue, please do not hesitate to contact Accuray Customer Support at 1. (877) 668-8667 ( USA) 1. (408) 716-4700 ( non USA )
Get a free legal consultation about this recall.
Talk to a Lawyer →Affiliate link
Products
- Brand
- Accuray Incorporated
- UPC
- Model number: PN 032384
Units Affected
30