FDA Device

Philips HeartStart Vehicle Wall Mount, Model M5528A with Philips HeartStart DC Power Module, Model M5529A. The HeartStart MRx portable external defibrillator has optional accessories which include a wall mount and a DC power module. The wall mount is intended for storage and transport of MRx devices in land based transport vehicles. The power module is an additional optional accessory intended to be connected to the wall mount as a means for charging MRx devices (batteries) when stored in the

Hazard

HeartStart Vehicle Wall Mount (Model M5528A) with DC Power Module (Model M5529A)--Incorrect wiring during installation by user may result in a smoke or fire hazard.

What You Should Do

Remedy

Philips Healthcare, Inc. notified customers by an "URGENT - Voluntary Medical Device Correction" letter, plus revised Instructions For Use (IFU) for the Vehicle Wall Mount accessory and the DC Power Module accessory of the HeartStart MRx which includes a revised description of correct installation and verification of performance. Customers were asked to follow the Actions in the customer/user section of the Urgent - Voluntary Medical Device Correction Notification. In this section, customers were given the instructions to correct their devices. Distribution of the Urgent - Voluntary Medical Device Correction Notification along with the IFU will be managed by Philips representatives in each affected geography. Direct questions or concerns about the recall to your local Philips representative.

Products

Philips HeartStart Vehicle Wall Mount, Model M5528A with Philips HeartStart DC Power Module, Model M5529A. The HeartStart MRx portable external defibrillator has optional accessories which include
Brand
Philips Healthcare Inc.
UPC
Units distributed by Philips Healthcare Inc. and shipped between September 1, 2004 and October 1, 2009.

Units Affected

3,265 units