FDA Device

Multi-Analyte Urine DAU Control (15 mL); Level 2, For in-vitro diagnostic use only. Lin-Zhi International, Inc. 670 Almanor Ave. Sunnyvale, CA 94085 Multi-Analyte Urine Drug of Drug Abuse (DAU) Controls are for use as controls in the quantitative and semi-quantitative calibration of Amphetamines Enzyme, Opiate Enzyme, Phencyclidine Enzyme, Cocaine Metabolite Enzyme, Benzodiazepine Enzyme, Barbiturate Enzyme, Methadone Enzyme and Propoxyphene Enzyme on a number of automated clinical analyzers.

Hazard

Customer complaint investigation found that the Multi-Analyte Urine DAU Level 2 control was incorrectly labeled and actually contained Multi-Analyte DAU Low Level Calibrators.

What You Should Do

Remedy

Lin Zhi International, Inc. sent a Recall notification letter dated August 5, 2011, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were asked to determine whether they wanted credit or a replacement product. The incorrectly labeled product should be disposed of properly. For questions regarding this recall call 408-732-3856.

Products

Multi-Analyte Urine DAU Control (15 mL); Level 2, For in-vitro diagnostic use only. Lin-Zhi International, Inc. 670 Almanor Ave. Sunnyvale, CA 94085 Multi-Analyte Urine Drug of Drug Abuse (DAU) Cont
Brand
Lin-Zhi International Inc
UPC
Catalog # 0808, lot number 1101137, Exp 1/28/2013

Units Affected

107 items kitted