ViroMax viral/bacterial filter. The filter reduces the risk of infection due to cross-contamination between patients For use on all patient populations in conjunction with other respiratory devices containing standard 15mm and/or 22mm fittings (such as breathing circuits) to filter respiratory gases where infection from airborne bacteria and viruses is a concern. Disposable, for single patient use only.
Hazard
Some ViroMax viral/bacterial filters are leaking at the point where the two halves of the filter are ultrasonically bonded together.
What You Should Do
A - M Systems sent a Urgent Notification of Product Recall/Correction dated March 31, 2010, to all affected customers. The letter identificed the product the problem and the action needed to be taken by the customer. Customers were instructed to cease use and/or distribution of the affected product and to immediately notify any of their sub-accounts with the same information. Customers are asked to provide A - M Systems with the amount distributed and amount remaining in inventory and to then destroy or discard any remaining product in their possession. Please call 360.683.8300, if you have any questions or concerns.
Get a free legal consultation about this recall.
Talk to a Lawyer →Affiliate link
Products
- Brand
- A M Systems Inc
- UPC
- Lot numbers: 551278, 551311, 551355, 551425, 551432, 551452, 551635, 551658, 551699, 551704, 551727, 551732, 551746, 551800, 551879, 551905, 551954 and 552095.
Units Affected
37800