Coherence AG Therapist Part number 5863506, equipped with a Coherence Therapist R2.0 or R2.1, and PRIMEVIEW 3iR2.0 or R2.1 Medical charged-particle radiation therapy system.
Hazard
Flat panel positioning calibration could be off by as much as 4 mm without the machine discovering detail.
What You Should Do
Remedy
Service calibration inspection/correction Update instructions TH038/07/S was issued on 11/06/2007 to all customers. The instructions identified the affected products, stated the reason for the update, and listed the steps to update the software. Customer notification was also issued. Software patch planned for release in June 2008.
Affected by this recall?
Get a free legal consultation about this recall.
Talk to a Lawyer →Affiliate link
Products
Coherence AG Therapist Part number 5863506, equipped with a Coherence Therapist R2.0 or R2.1, and PRIMEVIEW 3iR2.0 or R2.1
Medical charged-particle radiation therapy system.
- Brand
- Siemens Medical Solutions USA, Inc
- UPC
- Serial numbers associated with RTT2.0 and RTT2.1
Units Affected
14 units.