Accuracy Cyberknife Robotic Radiosurgery System, medical charged particle radiation therapy device. Model number 020700 (axum/standard treatment couch) and 021756 hand controller. The device is indicated for treatment planning and image-guided stereotactic radiosurgery and precision radiotherapy of lesions, tumors and conditions anywhere in the body when radiation treatment is indicated.
Hazard
Couch may move unexpectedly, which may result in patient impacting the linear accelerator.
What You Should Do
Remedy
All affected sites were notified via customer advisory notice. The Standard Treatment couch firmware will be updated and all affected sites will be retrofitted. Contact Customer Service at 1-877-668-8667 (USA) or 1-408-716-4700 (non-USA) or customersupport@accuray.com for additional information.
Affected by this recall?
Get a free legal consultation about this recall.
Talk to a Lawyer →Affiliate link
Products
Accuracy Cyberknife Robotic Radiosurgery System, medical charged particle radiation therapy device. Model number 020700 (axum/standard treatment couch) and 021756 hand controller. The device is indi
- Brand
- Accuray Inc