Prelude Short Sheath Introducer, Carolina Kidney and Endocrine - Declot tray, K12T-02693, Merit Medical Systems, Inc., South Jordan, Utah 84095. To provide access and facilitate the percutaneous introduction of various devices into veins and/or arteries while maintaining hemostasis for a variety of diagnostic and therapeutic procedures. This device can also provide access to a native or synthetic graft used for hemodialysis. The side port of the sheath allows adequate flow to perform temporar
Hazard
External tubing may separate during use, allowing blood loss.
What You Should Do
Remedy
All of Merit's US sales representatives and distributors were notified by Prelude Short Sheath Introducer - Product Recall letter on 4/14/2009. They were instructed to promptly contact their customers. Customers were instructed to isolate, remove from inventory, and return all affected inventory. A Product Retrieval Form was to be returned to Merit. Further information is available at 801-316-4998 or 801-208-4344.
Affected by this recall?
Get a free legal consultation about this recall.
Talk to a Lawyer →Affiliate link
Products
Prelude Short Sheath Introducer, Carolina Kidney and Endocrine - Declot tray, K12T-02693, Merit Medical Systems, Inc., South Jordan, Utah 84095.
To provide access and facilitate the percutaneous intr
- Brand
- Merit Medical Systems, Inc.
- UPC
- Lot number V686326.
Units Affected
9 units