VERTIER Surgical Table, Catalog #0788-100-000 (Standard table); distributed by Stryker Communications, Flower Mound, TX. Is an AC powered, general purpose, mobile surgical table with electro-hydraulically controlled movements which are part of a control feedback system.
Hazard
Hydraulic lines responsible for tilting the surgical table have the potential to be severed when articulating table to its lowest position, possibly resulting in the unexpected and rapid movement of the table.
What You Should Do
Remedy
Urgent: Device Recall letters were sent to consignees on 07/02/08 via FedEx with return receipt. Consignees were asked to discontinue use of the tables until a Stryker representative has come to the facility and updated the table. The firm has identified and validated replacement parts that prevent this occurrence.
Affected by this recall?
Get a free legal consultation about this recall.
Talk to a Lawyer →Affiliate link
Products
VERTIER Surgical Table, Catalog #0788-100-000 (Standard table); distributed by Stryker Communications, Flower Mound, TX.
Is an AC powered, general purpose, mobile surgical table with electro-hydrauli
- Brand
- Stryker Communications Corp
- UPC
- All lot codes/serial numbers.
Units Affected
52 units.