Siemens Coherence AG Therapist, Part No: 5863506, medical charged-particle radiation therapy system with software version 2.1. Siemens AG, Concord, CA.
Hazard
Image Orientation Incorrect (Image Shifts) - 1) if the reference image has been calibrated for centering and the reference image is used for patient positioning, the image will shift when the positioning tools are used. 2) This problem occurs only when non-square reference images are used, such as when images are cropped. At this point, the images will refresh, and the center of the reference im
What You Should Do
Remedy
On 10/03/2007, the firm sent Consignees a Customer Safety Advisory Notice informing them of the two potential issues with the display of reference images used for position correction. Consignees will be informed when a solution to this issue has been validated.
Affected by this recall?
Get a free legal consultation about this recall.
Talk to a Lawyer →Affiliate link
Products
Siemens Coherence AG Therapist, Part No: 5863506, medical charged-particle radiation therapy system with software version 2.1. Siemens AG, Concord, CA.
- Brand
- Siemens Medical Solutions USA, Inc
- UPC
- Serial Numbers: 84601, 84602, 85703, 85714, 85718, 85720, and 00E0815479E1
Units Affected
7 units