FDA Device

OSI, Allegro 6800 Imaging Table (Operating), single pedestal AC or battery operated table, Orthopedic Systems, Inc., Union City, CA

Hazard

Failure to anchor; due to a failure in the locking mechanism, the product may not anchor properly.

What You Should Do

Remedy

The firm issued a Product Correction Notification letter by mail on 02/28/2007 and follow up with phone calls. The letter notified users that a required no-cost upgrade of the removal and replacement of a microprocessor that controls the logic for the floor locking mechanism will be conducted by the recalling firm.

Products

OSI, Allegro 6800 Imaging Table (Operating), single pedestal AC or battery operated table, Orthopedic Systems, Inc., Union City, CA
Brand
Mizuho Orthopedic Systems Inc
UPC
Serial Numbers: 12-19, 21-61, 64-67, 70-71, 73-83, 85-93, 95, 97 and 99-107.

Units Affected

117 units