OSI, Allegro 6800 Imaging Table (Operating), single pedestal AC or battery operated table, Orthopedic Systems, Inc., Union City, CA
Hazard
Failure to anchor; due to a failure in the locking mechanism, the product may not anchor properly.
What You Should Do
Remedy
The firm issued a Product Correction Notification letter by mail on 02/28/2007 and follow up with phone calls. The letter notified users that a required no-cost upgrade of the removal and replacement of a microprocessor that controls the logic for the floor locking mechanism will be conducted by the recalling firm.
Affected by this recall?
Get a free legal consultation about this recall.
Talk to a Lawyer →Affiliate link
Products
OSI, Allegro 6800 Imaging Table (Operating), single pedestal AC or battery operated table, Orthopedic Systems, Inc., Union City, CA
- Brand
- Mizuho Orthopedic Systems Inc
- UPC
- Serial Numbers: 12-19, 21-61, 64-67, 70-71, 73-83, 85-93, 95, 97 and 99-107.
Units Affected
117 units