Heartport Direct Flow Arterial Cannula, product code DFK24. The device is sold separately and also in the following Kit Model Numbers: EDNN241, EDnn242, EDNN243, and EDNN244.
Hazard
During cardiopulmonary bypass (CPB) the distal tip of the arterial cannula can become disconnected from the body of the cannula.
What You Should Do
Remedy
The recall was initiated on 4/27/2006. The hospitals were notified via letters to the Director of Material Management, Director of the OR as well as the Director of the Cardiovascular Suite in the US. The letters were sent via certified mail delivery. For international, notification was made to the J&J affiliates (Gargrave and Italy) by the European Authorized Representative. Each country will follow local procedures for notitying customers.
Affected by this recall?
Get a free legal consultation about this recall.
Talk to a Lawyer →Affiliate link
Products
Heartport Direct Flow Arterial Cannula, product code DFK24. The device is sold separately and also in the following Kit Model Numbers: EDNN241, EDnn242, EDNN243, and EDNN244.
- Brand
- Ethicon, Inc. US
- UPC
- Product Code DFK24 Lot Numbers: MS0206031; MS0206032; MS0505014; MS0505020; MS0605009; MS0605038; MS0605039; MS0705034; MS0705035; MS0705044; MS0905040; MS0905066; MS1105041; MS1105061; MS1205043.
Units Affected
368