superDimension Bronchus System (AAS00016-01), an image-guided localization system used to guide endoscopic tools to predetermined targets within the pulmonary tract.

Hazard

The action is being taken in response to certain incidents that have been reported to the company, one of which concerned an adverse event in which a patient incurred a pneumothorax during a procedure in which a superDimension Bronchus System was used. The biopsy results for the patient were taken from a different area than the confirmed navigation site that was indicated by the system.

What You Should Do

Remedy

An Urgent Medical Device Field Action letter, dated 11/30/2006, was sent to medical facilities who have the superDimension Bronchus System installed. The letter infoms the facility to immediately stop using the Bronchus System, and move all inventory of the disposable system components to a controlled quarantine area. A superDimension representative will contact the consignees to schedule a time to retrieve the quarantined inventory, at which time a verification form will be completed.

Products

superDimension Bronchus System (AAS00016-01), an image-guided localization system used to guide endoscopic tools to predetermined targets within the pulmonary tract.
Brand
SuperDimension, Inc.
UPC
superDimension Bronchus System with the following serial numbers: S0030019, S030051, S030052, S030054, S030055, S030058 thru S030062, S030067, S030068, S030077 thru S030081; S030084, S030085, S030088 thru S030095; S030097 thru S030100, S030103, S030104, S030105, S030390, S030447, S030449, S0030009, S0030011, S0030013, S0030014, S0030015, S0030021, S0030022, S0030026, S0030042, S030043 thru S030050, S030053, S030056, S030057, S030063, S030064, S030066, S030069 thru S030076, S030082, S030086, S030087, S030096, S030102, S030477.

Units Affected

74 Bronchus Systems (37 US, 37 OUS)