Draeger TI500. Isolette Infant Incubator. Catalog Number: MU20505. Draeger Medical, Inc., Telford, PA 18969

Hazard

Heating failure- The power board which controls the incubator heater may not regulate the temperature properly resulting in a high temperature alarm and a loss of temperature control within the patient compartment.

What You Should Do

Remedy

The firm notified consignees on 2/1/08, by an Important Safety Notice. The letter instructed users to remove any units from use until a Draeger Service Representative can replace the power board. A Service Representative will be in contact to schedule a service visit. For questions concerning this recall please contact 215-660-2418 or Anbu.arasu@draeger.com.

Products

Draeger TI500. Isolette Infant Incubator. Catalog Number: MU20505. Draeger Medical, Inc., Telford, PA 18969
Brand
Draeger Medical, Inc.
UPC
Serial Numbers: PK05967, PK05990, PK05973, PK05986, PK05987, PK05988, PK05962, PK05963, PK05964, PK05965, PK05989, PK05976, PK05972, LT05951, PK05968, LT05952, PK05969, and PK05974.

Units Affected

18 units