Terumo Advanced Perfusion System 1, 6 inch diameter Roller Pump; Model 801041.
Hazard
The tube clamp mechanism on the roller pump may not open when the rotation knob is turned, making it difficult to remove and/or insert tubing.
What You Should Do
Remedy
The firm is visiting the consignees as complaints are received and making corrections. Corrections were made in August 2005 and April 2006 to date. Other units will be repaired if complaints are received and the firm issued a Device Correction Letter dated 6/6/07 notifying all users of the problem and appropriate action to take should it occur.
Affected by this recall?
Get a free legal consultation about this recall.
Talk to a Lawyer →Affiliate link
Products
Terumo Advanced Perfusion System 1, 6 inch diameter Roller Pump; Model 801041.
- Brand
- Terumo Cardiovascular Systems Corp
- UPC
- Serial number range 0033 through 1308 may be affected by this problem. Serial numbers 1060 and 1148 have been corrected.
Units Affected
1982 total for both models