Tucker Embryo Catheter, Stylet, For use with Embryo Transfer Catheter, 23 CM, Sterile, Single Use Only, Order #: ft r 223, Manufactured by Fertility Technology Resources, Inc., Marietta, GA 30062

Hazard

Expiration Date for Sterility; The 5 year dating could not be validated/supported for sterility assurance.

What You Should Do

Remedy

All customers were contacted by telephone starting 9/13/2007. They were asked to return all Tucker family of catheters and all Morton IUI catheters that are on the shelves. A UPS call tag was sent to each customer to facilitate the return of the products. A follow up Urgent Device Recall letter was sent to all customers on 10/9/2007 reminding them to return all outstanding inventory when the UPS call tag arrives.

Products

Tucker Embryo Catheter, Stylet, For use with Embryo Transfer Catheter, 23 CM, Sterile, Single Use Only, Order #: ft r 223, Manufactured by Fertility Technology Resources, Inc., Marietta, GA 30062
Brand
Fertility Technology Resources, Inc.
UPC
All units shipped before August 17, 2007

Units Affected

not reported