DynaRad Phantom Portable X-Ray System; Del Medical Imaging Corporation, 11550 West King Street, Franklin Park, IL 60131-1330; Phantom model (PH-150-CM & PH-150-G)

Hazard

The x-ray tube head assembly of the Phantom Portable X-Ray System may detach from the boom arm assembly. The tube head could fall and contact the patient or operator.

What You Should Do

Remedy

Del Medical sent Field Correction Recall letters dated 10/3/06 to the direct accounts, informing them of the potential for the shoulder bolt in the yoke assembly to come loose. If the shoulder bolt comes loose, the tube head could potentially fall and contact the patient or operator. The accounts were requested to locate the affected systems and take them out of service until they are inspected and, if needed, repaired.

Products

DynaRad Phantom Portable X-Ray System; Del Medical Imaging Corporation, 11550 West King Street, Franklin Park, IL 60131-1330; Phantom model (PH-150-CM & PH-150-G)
Brand
Del Medical Systems Group
UPC
Generator serial numbers: 03460-1003, 03477-1003, 03499-1003, 03550-1003, 03647-1003, 03704-1003, 03734-1003, 03744-1003, 03759-1003, 03765-1003, 03823-1003, 03938-1103, 03941-1103, 03953-1003, 04227-1103, 04282-1203, 04303-1203, 04330-1203, 04477-1203, 04497-1203, 04666-1203, 04704-1203, 00013-0104, 00035-0104, 00107-0104, 00122-0104, 00127-0104, 00188-0104, 00192-0104, 00332-0104, 00336-0104, 00405-0104, 00417-0104, 00430-0104, 00474-0104, 00545-0204, 00555-0204, 00568-0204, 00709-0204, 00719-0204, 00741-0204, 00754-0204, 00776-0204, 00995-0304, 00999-0304, 01101-0304, 01291-0304, 01295-0304, 01369-0304, 01423-0404, 01450-0404, 01468-0404, 01473-0404

Units Affected

53 units