Medication Bottle (U22-3), a component of the MicroAir Vibrating Mesh Nebulizer, Model NE-U22V/NE-22VAC; Distributed by: Omron Healthcare, Inc., 300 Lakeview Parkway, Vernon Hills, IL 60061
Hazard
A potential failure of the vibrating mechanism of the medication bottle which causes a failure to transmit vibration and results in no nebulization.
What You Should Do
Omron sent 'Removal Notices' dated 11/2/04 to all of their direct accounts receiving the nebulizers during the affected time frame. The accounts were informed of the potential failure of the vibrating mechanism, and listed the serial numbers of the nebulizers that may contain the affected medication bottles. Medication bottles with either a blue dot near the bottle lever or a blue O-ring around the body of the bottle were manufactured after corrections were made and are not subject to the recall. Affected product should be returned to Omron for replacement. Any questions were directed to Customer Service at 800-231-4030.
Get a free legal consultation about this recall.
Talk to a Lawyer →Affiliate link
Products
- Brand
- Omron Healthcare, Inc.
- UPC
- Bottle lots 43A, 53A, 63A, 73A, 83A, 93A, X3A, Y3A and Z3A, a component of Nebulizer model NE-U22 serial numbers 3400001A and higher, 3500001A and higher, 3600001A and higher, 3700001A and higher, 3800001A and higher, 3900001A and higher, 3X00001A and higher, 3Y00001A and higher, and 3Z00001A through 3Z00696A. Medication bottles with either a blue dot or a blue O-ring were manufactured after corrections were made and are not subject to the recall.
Units Affected
63,560 units