Medical Air Compressor Model #8413419

Hazard

Compressors not providing sufficient supply pressure to connected medical ventilator

What You Should Do

Remedy

The recalling firm issued a Medical Device Recall Letter dated 1/22/07 to their customers via Certified Mail. An authorized service representative will inspect each potentially affected device. If a device is found to have an affected dryer assembly, it will be repaired free of charge.

Products

Medical Air Compressor Model #8413419
Brand
Draeger Medical, Inc.
UPC
Model number 8413419. Catalog number 8413419. Serial Numbers ARXC-001; ARXC-0002; ARXC-0009; ARXC-0011; ARXC-0012; ARXC-0013; ARXC-0014; ARXC-0015; ARXC-0016; ARXC-0017; ARXC-0019; ARXC-0020; ARXD-0034; ARXD-0035; ARXD-0036; ARXD-0037; ARXD-0038; ARXD-0039; ARXD-0040; ARXE-0081; ARXE-0082; ARXE-0083; ARXE-0084; ARXE-0085; ARXE-0086; ARXE-0087; ARXE-0088; ARXE-0089; ARXE-0090; ARXE-0091; ARXE-0092; ARXE-0093; ARXE-0094; ARXE-0095; ARXE-0096; ARXE-0097; ARXE-0098; ARXE-0099; ARXE0100; ARXE-0101; ARXE-0102; ARXE-0103; ARXE-0104; ARXE-0106; ARXE-0107; ARXE-0108; ARXE-0109; ARXE-0110; ARXE-0111; ARXE-0112; ARXE-0113; ARXE-0114; ARXE-0115; ARXE-0116; ARXE-0117; ARXE-0118; ARXE-0119; ARXE-0120; ARXE-0121; ARXE-0122; ARXE-0123; ARXE-0124; ARXE-0125; ARXE-0126; ARXE-0127; ARXE-0128; and ARXE-0129.

Units Affected

67 units