Spectrum Patient Monitor, Physiological, Patient with arrhythmia detection or alamrs. Monitor can monitor, display, trend and print a patient''s physiological parameters. The device has a 12.1 inch color display and has a standard configuration of a 3 or 5 lead ECG, Masimo SET SpO2, non-invasive blood pressure, respiration, continuous temperature and IV Drug Calculations. Optional digital displays are provided for invasive blood pressure (up to four) anesthetic agents, O2 and NO2, ST, and CO2

Hazard

Software anomalies which include shut down while printing, invalid diastolic blood pressure display, and incomplete and inaccurate interpretation reports of 12-Lead ECG data.

What You Should Do

Remedy

Datascope sent out a Field Correction notice on 9/7/2005 to all Hospital administrators who have the monitors. This was sent out via certified mail, return receipt requested. A Datascope Service Rep will conduct and document the corrective action during a visit to the customer.

Products

Spectrum Patient Monitor, Physiological, Patient with arrhythmia detection or alamrs. Monitor can monitor, display, trend and print a patient''s physiological parameters. The device has a 12.1 inch
Brand
Datascope Corp
UPC
All Spectrum Monitors that have the following Software Versions: A.19, A.20, B.23, B.24, B.25, B.27, C.18, D.07, E.07, F.09, F.16, F.17, F.26.

Units Affected

2318 monitors