FDA Device

Medtronic Activa Deep Brain Stimulation (DBS) system including the following system components: Model 7424 Itrel II Neurostimulator (DBS applications only); Model 7426 Soletra Neurostimulator; Model 7428 Kinetra Neurostimulator; Models 3387, 3389 DBS Leads; and Models 7482,7495 DBS Extensions. Medtronic, Inc. 710 Medtronic Parkway, Minneapolis, MN 55432-5604, USA. The Medtronic Activa System is an implantable, multiprogrammable system that delivers electrical stimulation to selected areas o

Hazard

A safety concern exists regarding the Medtronic Activa Deep Brain Stimulation system instructions for performing MRI (magnetic resonance imaging). Medtronic's instructions have until now recommended a use of a displayed MRI head specific absorption rate (SAR) of not more than 0.4 W/kg. The recommended maximum displayed head SAR is now 0.1 W/kg.

What You Should Do

Remedy

The firm''s recall consists of providing revised instructions for the Activa DBS. The revision modifies the recommended SAR to use during MRI of patients who have an Activa DBS. Use of the Activa DBS is otherwise unaffected by this recall. A Dear Healthcare Provider letter and MRI Radiologist letter were sent on 11/30/05 and 12/05/05 to provide this information.

Products

Medtronic Activa Deep Brain Stimulation (DBS) system including the following system components: Model 7424 Itrel II Neurostimulator (DBS applications only); Model 7426 Soletra Neurostimulator; Model
Brand
Medtronic Inc, Neurological & Spinal Division
UPC
All serial numbers and Lot codes are affected.