FDA Device

IMPAX 4.5 Systems, used in the acceptance, transfer, display, storage and digital processing of medical images.

Hazard

Possible corrupted image appearing after System Start.

What You Should Do

Remedy

On 06/30/2006 Agfa upgraded the Software on the installed units using a Mandatory Service Bulletin DD+DIS225.06E.

Products

IMPAX 4.5 Systems, used in the acceptance, transfer, display, storage and digital processing of medical images.
Brand
AGFA Corp.
UPC
All Impax 4.5 sites running Solaris 9 (Sun OS 5.9) that do not have Sun Patch ID 118305-08 installed.

Units Affected

2 units