ARCHITECT B12 Reagent, list 06C09-22 (4 x 100 tests), 06C09-27 (1 x 100 tests); an in vitro diagnostic kit consisting of microparticles, conjugate, assay diluent, pre-treatment 1, pre-treatment 2, and pre-treatment 3; Abbott Laboratories, Abbott Park, IL 60064
Hazard
The 30-day onboard storage information is not included in the barcode labels for these two lots. As a result, the ARCHITECT system software is unable to track how long a reagent kit has been stored onboard the ARCHITECT instrument.
What You Should Do
Abbott affiliates in Canada, Germany and Australia were e-mailed copies of the recall letter on 3/8/06 for translation where needed and dissemination to their accounts receiving the suspect lots. The end users were advised that the 30-day onboard storage information is not included in the barcode for the two lots. As a result the ARCHITECT system software is unable to track how long the reagent kit has been stored onboard. The accounts were instructed to manually track the onboard storage time for the two lots, and discard the kits after 30 days of onboard storage.
Get a free legal consultation about this recall.
Talk to a Lawyer →Affiliate link
Products
- Brand
- Abbott Laboratories MPG
- UPC
- lot numbers 33079M200, 33079M201
Units Affected
135 kits