COBAS TaqMan Analyzer
Hazard
Improper seating of a fiber optic cable which may cause pixel cross-talk leading to falsely elevated test results or cases of misalignment to sample-to-result mismatch.
What You Should Do
Remedy
An Urgent Product Recall letter was sent to all customers with a COBAS TaqMan 96 and 48 analyzer on 5/24/2004 by overnight delivery with delivery confirmation. On May 25, 2004, All Roche Field Service Engineers were instructed to inspect the analyzers to confirm that the cables were properly seated.
Affected by this recall?
Get a free legal consultation about this recall.
Talk to a Lawyer →Affiliate link
Products
COBAS TaqMan Analyzer
- Brand
- Roche Molecular Systems Inc
- UPC
- COBAS TaqMan Analyzer Serial Numbers: 381046; 381053; 391159; 391166; 391181; 371036; 391130; 391095; 381052; 391191; 391096; 381044; 391131; 381050; 381054; 391180; 381045; 391177; 391189; 381047; 381051; 381055; 391182; 391157; 391148; 391135; 391143; 391141; 391138; 391175; 391144; 391161; 391150; 391156.
Units Affected
34