Roche brand AccuData GTS Plus System; Catalog number 3000249, material number 03000249001.
Hazard
If foil wrapped packages or other metallic objects are lodged behind the supply drawer, an electrical short may result and high heat generated; this could result in plastic melting and battery leakage for customers using 'C' cell batteries to power their base and could render the system inoperable (system won't turn on) for customers using the A/C adaptor without batteries.
What You Should Do
Remedy
Consignees were notified via letter dated 4/18/05, which included instructions on how to inspect their unit for damage, and how to prevent future damage.
Affected by this recall?
Get a free legal consultation about this recall.
Talk to a Lawyer →Affiliate link
Products
Roche brand AccuData GTS Plus System; Catalog number 3000249, material number 03000249001.
- Brand
- Roche Diagnostics Corp.
- UPC
- All units.