FDA Device

Digene Hybrid Capture System CMV DNA Test (version 2.0), packaged in a cardboard box containing test reagents for performing 60 tests. Digene Catalog #4403-2003.

Hazard

In vitro diagnostic test kit may produce false negative clinical results.

What You Should Do

Remedy

The recalling firm notified consignees by phone 5/28/04 and by faxed letter 6/2/04. The letter advised of possible false negative test results and that such results should be retested immediately with replacement kits provided by the firm. The notification further advised that positive results should be considered valid if the assay meets all verification and calibration criteria. Consignees were requested to return product and response form.

Products

Digene Hybrid Capture System CMV DNA Test (version 2.0), packaged in a cardboard box containing test reagents for performing 60 tests. Digene Catalog #4403-2003.
Brand
Digene Corp
UPC
Lot number 6234, EXP 10/20/04

Units Affected

206 kits