Roche brand OMNI 7 clinical chemistry analyzer; catalog numbers GD0435 and GD0435R.
Hazard
If the sample volume is insufficient, the analyzer may issue erroneous results without any indication to the user that the problem is insufficient sample volume.
What You Should Do
Remedy
Consignees were notified by letter mailed on or about December 15, 2003. Customers are instructed to assure an adequate sample volume until new software becomes available.
Affected by this recall?
Get a free legal consultation about this recall.
Talk to a Lawyer →Affiliate link
Products
Roche brand OMNI 7 clinical chemistry analyzer; catalog numbers GD0435 and GD0435R.
- Brand
- Roche Diagnostics Corp.
- UPC
- All units with software version 5.61 or below.