IRMA SL Blood Analysis System CC pH, pCO2, pO2, Hct, Na+, K+, iCa Cartridge, Part Number 039903(M3586A). There are 25 disposable cartridges in a box of IRMA SL Blood Analysis System CC pH, pCO2, pO2, Hct, Na+, K+, iCa Cartridges, Part Number 039903(M3586A).
Hazard
An incorrect lot designation, when entered into the blood analysis device, caused the device to give an error message and not operate.
What You Should Do
Remedy
The consignees were notified of the recall by telephone on July 14, 2003 and requested to return the product.
Affected by this recall?
Get a free legal consultation about this recall.
Talk to a Lawyer →Affiliate link
Products
IRMA SL Blood Analysis System CC pH, pCO2, pO2, Hct, Na+, K+, iCa Cartridge, Part Number 039903(M3586A). There are 25 disposable cartridges in a box of IRMA SL Blood Analysis System CC pH, pCO2, pO2,
- Brand
- Diametrics Medical, Inc.
- UPC
- The lot on the boxes of the product is ALHCB (this is the correct lot) and is incorrectly printed as ALOHC on the packages of the recalled disposable cartridges.
Units Affected
103 boxes of 25 cartridges each.