MagNA Pure LC Instrument; Catalog number 2236931.
Hazard
A software bug in version 3.0 may result in prolonged mixing time, evaporation of elution volume and biased sample results for various protocols.
What You Should Do
Remedy
An urgent product correction letter dated 6/26/03 was sent to each customer advising them not to use DNA protocols until the software update has been installed.
Affected by this recall?
Get a free legal consultation about this recall.
Talk to a Lawyer →Affiliate link
Products
MagNA Pure LC Instrument; Catalog number 2236931.
- Brand
- Roche Diagnostics Corp.
- UPC
- All systems using software version 3.0
Units Affected
17