Clinitron At Home Air Fluidized Therapy Unit
Hazard
Power cord overheating
What You Should Do
Consignees were notified by letter sent January 14, 2002 warning them of the hazard involved with the defective part. They were informed that Hill Rom needed to replace the power cord on each unit to correct the possibility of the power cord from overheating, melting or smoking. The maintenance would be performed at the location of each unit (most of which are at their rental facilities located nationwide) by the firm''s servicing representatives. Meanwhile, customers were urged to check the power cord to ensure that it is fully inserted into the wall outlet and the connector on the Clinitron At Home Air Fluidized Therapy Unit. The modification was to be completed by August 31, 2002.
Get a free legal consultation about this recall.
Talk to a Lawyer →Affiliate link
Products
- Brand
- Hill-Rom Manufacturing, Inc.
- UPC
- Model 206 (CS), Serial numbers 101671 and 103443 510K: K942184
Units Affected
1703 units