Free T4 Control Pack, list 9C03-10; the kit consists of 3 bottles of controls, L - 0.7, M - 1.2 and H - 3.0; Abbott Laboratories, Abbott Park, IL 60064
Hazard
The Free T4 Controls, may yield control values that are high and outside of the package insert ranges.
What You Should Do
The Free T4 Control Pack, list 9C03-10, and the IMx Free T4 Reagent Kit, list 2222-20, were recalled by letter dated 5/30/03. The accounts were informed of the out of range control values, and the potential for depressed patient values with the IMx kit, and were requested to discontinue use and destroy the affected lots of product. Abbott sent a Device Correction letter dated 6/4/03 to those accounts receiving the Architect Free T4 Calibrator kits, informing them of the problems experienced with the The Free T4 Control Pack, list 9C03-10, and the IMx Free T4 Reagent Kit, list 2222-20, made from the same raw material. The accounts were informed that the Architect Calibrator was performing as expected, but as a precaution recommended conducting a daily QC with non-Abbott Controls and to report back to Abbott any out of range non-Abbott control results.
Get a free legal consultation about this recall.
Talk to a Lawyer →Affiliate link
Products
- Brand
- Abbott Laboratories HPD/ADD
- UPC
- List 9C03-10, lot numbers 95881Q100 and 95881Q101
Units Affected
3247 kits